UNIVERSAL BREAKDOWN INSTALLATION UPU-10

Delivery time 15 working days
Download Guide  Руководство в pdf для UNIVERSAL BREAKDOWN INSTALLATION UPU-10
Order consultation
appointment
Specifications
Equipment

Output voltage setting limits, kV, - 1; 3; 10
Reduced relative error of setting and measuring output
   voltage,%, no more, - 3
Maximum output current, mA, - 100
Reduced relative error of current measurement,%, no more - 3     
Threshold of current protection, mA, - 102 ± 2
Power consumption , V.A, no more than 1000
Overall dimensions, mm, control unit - 320x200x355

high voltage unit - 315х290х410 
Weight, kg, no more than: 
control unit - 15
high voltage unit - 25                                                                                 
Operating time in continuous mode at maximum power, hour, - 8
Average product service life, years - 6
NOTE. The kilovoltmeter of the installation measures on an alternating voltage - the rms value, on a rectified voltage - the amplitude value of the voltage. 
An AC milliammeter measures the rms value, at the rectified
average ( rectified  average) current value.

Name

amount

1

Universal breakdown installation UPU-10M.

1

2

Network wire.

1

3

High-voltage connecting wire.

1

4

Ground wire.

2

5

Universal breakdown installation UPU-10M. Manual. UPU-10M / 00.00.00.00RE.

1

6

Universal breakdown installation UPU-10M.

Passport UPU-10M / 00.00.00.00PS.

1

Universal breakdown installation UPU-10M (hereinafter referred to as the Installation) is intended for testing the insulation of electrical equipment and materials with alternating sinusoidal voltage with a frequency of 50 Hz and rectified voltage of negative polarity, adjustable within 0-10 kV with an output current of up to 100 mA.  
The field of application is electrical engineering and power engineering.
Due to the increased output current, the installation can be used for testing the insulation of large electrical machines at alternating voltage. It can be successfully used for testing protective equipment for appropriate voltages (gloves, etc.), as well as for charging capacitive energy storage devices.
The unit is designed for operation indoors or under a canopy at operating air temperatures from plus 5 ° C to plus 40 ° C, relative humidity up to 80% at a temperature of plus 20 ° C and atmospheric pressure 84.0 - 106.7 kPa (630 - 800 mm Hg).
Power supply - single-phase sinusoidal alternating current network with a voltage of 220-20 V, a frequency of 50-1 Hz. 
The installation is serviced by one operator with the right to work on installations with voltages above 1000V.